| 离子色谱法测定水中溴离子与碘离子 |
| Determination of Bromide and Iodide in Water by Ion Chromatography |
| |
| DOI: |
| 中文关键词: 溴离子 碘离子 离子色谱法 水质 |
| 英文关键词: Bromide Iodide Ion chromatography Water quality |
| 基金项目:国家水体污染控制与治理科技重大专项基金“水环境质量监测技术方法研究课题”资助项目(2009ZX07527-001) |
|
| 摘要点击次数: 2210 |
| 全文下载次数: 3095 |
| 中文摘要: |
| 采用离子色谱法测定水中溴离子与碘离子,选择AS19阴离子交换分离柱、KOH淋洗液自动发生器、抑制型电导检测。溴离子与碘离子分别在14.8 μg/L~100 mg/L与30.4 μg/L~100 mg/L范围内线性良好,检出限分别为3.7 μg/L与7.6 μg/L,环境水样平行测定的RSD分别为0.2%~1.5%与0.3%~1.6%,两个质量浓度水平的加标回收率分别为91.8%~105%与89.5%~93.4%。 |
| 英文摘要: |
| An ion chromatography method for determining Br- and I- in water was studied.The separation of the anions was achieved on an IonPac AS19 Column with KOH as eluent,and the detection was performed by a conductivity detection mode.Good linearity were obtained within the concentration range of 14.8 μg/L~100 mg/L for Br- and 30.4 μg/L~100 mg/L for I-.The method detection limits of Br- and I- were 3.7 μg/L and 7.6 μg/L,respectively.The RSD of the samples were 0.2%~1.5% (Br-) and 0.3%~1.6% (I-).The recoveries of Br- in water samples were 91.8%~105%,and the recoveries of I- were 89.5%~93.4%. |
|
查看全文
查看/发表评论 下载PDF阅读器 |
| 关闭 |