ICP-MS碰撞池技术消除复杂水样中硒的基体干扰 |
Elimination of Matrix Interference for Determining Selenium in River Sample by Inductively Coupled Plasma Mass Spectrometry KED Technique |
|
DOI: |
中文关键词: 硒 电感耦合等离子体质谱法 碰撞池技术 基体干扰 水质 |
英文关键词: Se ICP MS KED technique Matrix interference Water quality |
基金项目: |
|
摘要点击次数: 1387 |
全文下载次数: 0 |
中文摘要: |
通过测定深圳某河流水样中的主要成分,确定ICP-MS碰撞模式测定Se的主要基体干扰,并对碰撞气流量进行条件优化和水样基体干扰消除试验,确定当氦气流量为3.0 L/min时,测定77Se的干扰能基本被消除。该方法测定实际水样的结果与原子荧光法的测定值无显著差异。 |
英文摘要: |
Components and content of the complex river samples in Shenzhen were studied. Major interference on determination of Se with ICP MS was confirmed. To eliminate the interference, ICP MS KED mode equipped with He as collision gas was used to examinate 77Se and 82Se. The flow rate of He was optimized, and a flow of 3.0 L/min of He was found as a suitable rate for fully eliminating interference on 77Se. But interference on 82Se can only be remove partly. Analysis results of Se in the river sample (77Se was selected) matched up the results determined by atomic flourescence spectrometry. |
查看全文
查看/发表评论 下载PDF阅读器 |
关闭 |