025-83701931 设为首页 加入收藏
文章摘要
影响顶空-气相色谱法测定水中吡啶准确度的关键因素探讨
Discussion on Accuracy in Determination of Pyridine in Water by Headspace Gas Chromatography
  
DOI:
中文关键词: 吡啶  顶空进样  气相色谱法  准确度  影响因素  水质
英文关键词: Pyridine  Headspace sampling  Gas chromatography  Accuracy  Influence factor  Water quality
基金项目:
作者单位
李江 张家港市环境监测站 
陈华 张家港市环境监测站 
苏慕珂 张家港市环境监测站 
摘要点击次数: 1023
全文下载次数: 0
中文摘要:
      通过试验,讨论盐析效应、顶空的平衡温度和平衡时间、水样的pH值、色谱条件等对顶空-气相色谱法测定环境水体中吡啶准确度的影响。优化上述条件后,方法在0 mg/L~3.95 mg/L范围内线性良好,检出限可达0.016 mg/L,测定下限为0.064 mg/L。用该方法对地表水和生活污水做3个质量浓度水平的加标回收试验,测定6次结果的RSD为 1.7%~5.7%,回收率为93.9%~113%。
英文摘要:
      Issues like salting out effect, headspace equilibrium temperature and time, pH value of the sample and GC parameters were discussed in determination of pyridine in water by headspace gas chromatography. The results showed that under the optimized parameters, the calibration curve of pyridine had a good linearity between the concentration of 0 mg/L and 3.95 mg/L, and the MDL was 0.016 mg/L. RSDs ranged from 1.7% to 5.7% by 6 duplicate tests. The spiked recoveries ranged from 93.9% to 113%.
查看全文   查看/发表评论  下载PDF阅读器