XAD-7吸附-毛细管柱气相色谱法测定工作场所空气中乙二醇 |
Determination of Ethylene Glycol by Using XAD 7 Adsorption Capillary Column Gas Chromatography in Workplace Air |
|
DOI: |
中文关键词: 乙二醇 离子交换树脂 气相色谱法 工作场所空气 |
英文关键词: Ethylene glycol Ion exchange resin Gas chromatography Workplace air |
基金项目:国家自然科学基金资助项目(31570515);苏州市科技支撑计划基金资助项目(SS201421,SS201523);苏州市环保科技基金资助项目(2015-3) |
|
摘要点击次数: 1275 |
全文下载次数: 0 |
中文摘要: |
采用离子交换树脂XAD-7吸附-毛细管柱气相色谱法测定工作场所空气中乙二醇,结果表明,该方法在0 mg/L~1 116 mg/L范围内线性良好,方法检出限为0.88 mg/L,当采样体积为7.5 L时,最低检出质量浓度为0.11 mg/m3。乙二醇标准溶液测定结果的RSD为1.3%,空白加标回收率为80.1%~107%,方法的采集效率为98.0%。 |
英文摘要: |
The method of determining ethylene glycol in the air of workplace was established by ion exchange resin XAD 7 adsorption capillary column gas chromatography. Experimental results showed good linear relationship in the range of 0 mg/L~1 116 mg/L and the lowest detection limit was 0.88 mg/L. The minimum detectable concentration was 0.11 mg/m3 when the sampling volume was 7.5 L. The relative standard deviation(RSD) of standard solution of ethylene glycol was 1.3%. The blank standard addition recoreries were between 80.1%and 107%, the collection efficiency was 98.0%. |
查看全文
查看/发表评论 下载PDF阅读器 |
关闭 |