025-83701931 设为首页 加入收藏
文章摘要
聚氨酯泡塑预富集-石墨炉原子吸收法测定水中铊
Determination of Thallium in Water by Polyurethane Foam Preconcentration Graphite Furnace Atomic Absorption Spectrometry
  
DOI:
中文关键词:   聚氨酯泡塑预富集  石墨炉原子吸收光谱法  水质
英文关键词: Thallium  Polyurethane foam preconcentration  GFAAS  Water quality
基金项目:国家环境保护标准制修订基金资助项目(2014-19);江苏省环境监测科研基金资助项目(1102)
作者单位
杜青 江苏省南京环境监测中心 
摘要点击次数: 586
全文下载次数: 46
中文摘要:
      水样经硝酸和过氧化氢消解,Tl+被氧化为Tl3+,在王水和Fe3+介质中,TlCl-4被聚氨酯泡塑吸附富集,沸水脱附后用石墨炉原子吸收光谱法测定。用硝酸钯-抗坏血酸基体改进剂提高测定灵敏度,方法在0 μg/L~250 μg/L范围内线性良好,方法检出限为001 μg/L。实际样品加标后6次测定结果的〖JP3〗RSD为30%~110%,平均加标回收率为950%~〖JP〗998%。有证标准物质的测定结果在标准值范围内,相对误差为-48%。
英文摘要:
      A method for determining thallium in water was developed by graphite furnace atomic absorption spectrometry. Tl+ was oxidized to Tl3+ in nitrate acid and hydrogen peroxide digestion, TlCl-4 was concentrated by polyurethane foam in aqua regia and Fe3+ solution, then desorbed in boiling water. Palladium nitrate and ascorbic acid were screened as matrix modifier for improving sensitivity. The method had a good linearity in the range of 0 μg/L~25.0 μg/L. The MDL was 0.01 μg/L. The RSDs of spiked sample measurements repeated 6 times were from 3.0% to 11.0%, and the average recoveries were between 95.0% and 99.8%. The measuring result of a certified reference material was within the prescribed limit, the RE was -4.8%.
查看全文   查看/发表评论  下载PDF阅读器
关闭