ICP- AES法测定防腐处理后木材中可溶性铜铬砷 |
Determination ofSoluble Copper, Chrom ium and Arsenic in Preservative TreatedW ood by ICP - AES |
|
DOI: |
中文关键词: 铜 铬 砷 电感耦合等离子发射光谱法 木材 防腐处理 |
英文关键词: Copper Chromium Arsenic ICP- AES W ood Preservativetreatment |
基金项目: |
|
摘要点击次数: 1545 |
全文下载次数: 1487 |
中文摘要: |
采用电感耦合等离子发射光谱法同时测定防腐处理后 木材中 可溶性铜、铬、砷, 优化了试验条件, 确定了最佳分析线。方法在 0.500 m g/L~ 10.0 mg/L 范围内线性良好, 铜、铬、砷的检出限分别为 0.003 m g/L 、0. 01 mg/L 、 0.07 m g/L, 木材样品平行测定的RSD为 0.5% ~ 1.2%, 加标回收率为 97. 0% ~ 102%, 与原子吸收光谱法的测定结果相吻合。 |
英文摘要: |
A methodwasestablishedtotestsoluble copper, chromium and arsenicin preservativetreated woodbyinductivelycoupledplasmaatomicemissionspectrom etry. Theexperimentalconditionswereoptimizedto producefollowingresults. Thebestanalyticallinesofspectra wereconfirmed. A goodlinearity wasinrangefrom0.500 mg/L to 10. 0 mg/L. The detection limits ofCu, Crand As were 0.003 mg/L, 0. 01 mg/L and0.07 mg/L respectively. The RSD ofparallelwoodsamplestestsrange wasfrom 0.5% to1.2%. Spikerecovery rangewasfrom 97.0% to102%. Theresultswerein agreem entwiththatofAAS. |